First Experimental Result

Our 10 cm² zero-gap alkaline electrolysis cell uses nickel electrodes and a laboratory-prepared porous separator, without PGM catalysts.

At a current density of 1 A/cm², the measured cell voltages were:

  • Room temperature: 2.20 V
  • 70 °C: 2.05 V
  • 80 °C: 1.98 V

These initial results were achieved before systematic catalyst optimisation. Further development will address higher current densities, long-term durability, gas purity, dynamic operation and system scale-up.

Engineering first. Catalysis next.

Validation and System Scale-Up

The results obtained so far provide a foundation for further development of our laboratory-scale electrolysis cell. Moving towards industrial applications requires additional testing and systematic performance evaluation.

Our next development steps include:

  • Long-term durability: monitoring cell voltage and degradation of electrodes and separators during extended operation.
  • Higher current densities: evaluating cell performance under industrially relevant operating conditions.
  • Efficiency and gas quality: measuring electrical energy consumption, Faradaic efficiency and gas crossover.
  • Dynamic operation: investigating the response to varying electrical power and opportunities for electrode conditioning and regeneration.
  • System scale-up: developing larger cells, assembling cell stacks and integrating balance-of-plant components.

Our goal is to connect laboratory research with a Slovenian Electrolyzer demonstration system through measurable and reproducible experimental results.